Description
DESCRIPTION: Bovie Surgi Center Pro Features
- The Bovie Surgi Center Pro is a 200-watt high-power operating room generator designed to deliver premium performance at significant cost savings to medical facilities.
- Incorporates 12 monopolar and bipolar RF energy modalities for precise surgical interventions.
- Offers Pure Cut, four modes of Blend, three modes of Coagulation, Bipolar Coagulation, and Bovie FCFS™ technology for seamless transition between Monopolar and Bipolar modes.
- Features enough power to perform procedures in surgery centers and most hospital operating rooms.
- Built with the highest standards of safety and quality for maximum reliability and economy.
- Includes specialized Laparoscopic Mode and Dual Monopolar Spray Activation for enhanced versatility and functionality.
SPECIFICATIONS: Bovie Surgi Center Pro Specifications
- Dimensions:
- Height: 6.5 In (16.5 cm)
- Width: 14.75 In (37.5 cm)
- Depth: 18.1 In (46 cm)
- Weight: < 20 lbs (< 9.07 kg)
- Operating Parameters:
- Ambient temperature range: 10° to 40° C
- Relative humidity: 30% to 75%, non-condensing
- Atmospheric pressure: 70kPa to 106kPa
- Warm-up time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.
- Transport and Storage:
- Ambient temperature range: -40° to +70° C
- Relative humidity: 10% to 100%, including condensation
- Atmospheric pressure: 50kPa to 106kPa
- General:
- Audible/Visible Output Meter: Yes
- Contact Monitoring Systems: BovieNEM
- Cooling System: Passive Cooling
- Technology: Isolated RF Output
- Pre-set Feature: Yes
- Monopolar HP Ports: 2
- Dual Activation Tone Alert: Yes
- Connector Ports/Power Switch: White
- Bipolar:
- Macro Bipolar: Watts: 80, VPeak Max Voltage: 600V, Frequency: 490 kHz ± 4.9 kHz
- Micro Bipolar: Watts: 80, VPeak Max Voltage: 500V, Frequency: 490 kHz ± 4.9 kHz
- Standard Bipolar: Watts: 50, VPeak Max Voltage: 250V, Frequency: 490 kHz ± 4.9 kHz
- Monopolar:
- Cut I: Watts: 200, VPeak Max Voltage: 1000V, Frequency: 490 kHz ± 4.9 kHz
- Cut II: Watts: 200, VPeak Max Voltage: 750V, Frequency: 490 kHz ± 4.9 kHz
- Blend (1) to Blend (4): Watts: 200, Frequency: 490 kHz ± 4.9 kHz
- Pinpoint Coagulation: Watts: 120, VPeak Max Voltage: 1800V, Frequency: 490 kHz ± 4.9 kHz
- Spray Coagulation: Watts: 120, VPeak Max Voltage: 4000V, Frequency: 350 to 450 kHz
- Gentle Coagulation: Watts: 120, VPeak Max Voltage: 450V, Frequency: 490 kHz ± 4.9 kHz