Biolight BLT V6

Single Product Request Form

Description

Features:

  • Patient Monitor and Vital Signs monitor.
  • Provides pulse rate, non-invasive blood pressure (NIBP), SpO2, CO2, and temperature monitoring.
  • Can store up to a thousand individual pieces of recorded data.
  • Weighs only 5.5 pounds, facilitating easy transport to patients.
  • Runs on a rechargeable lithium-ion battery with 8 hours of continuous use.
  • Equipped with a 7-inch color LCD display with anti-glare for easy reading of data.
  • Suitable for various settings including ambulance, hospital, intra-hospital transport, and emergency.

Specifications:

  • Dimensions:
    • Height: 5” (13 cm)
    • Width: 6.3” (16 cm)
    • Depth: 10.2” (26 cm)
    • Weight: 5.5 lbs. (2.5 kg)
  • Display:
    • Type: Color TFT LCD
    • Size (Diagonal): 7” (17.8 cm)
    • Resolution: 234×480 pixels
  • Recorder:
    • Type: Thermal dot array
    • Paper Width: 50 mm
    • Recording Width: 40 mm
    • Recording Speed: 25 mm/s
    • Recording Waveform: Double tracks
  • Battery:
    • Type: Rechargeable lithium-ion battery
    • Size: 10 cm × 7.8 cm × 2 cm
    • Weight: <360 g
    • Rated Voltage: 11.1 VDC
    • Capability: 4000 mAh
    • Operating time: 8 hours
    • Charge Time: 6 hours to 100% (Standby)
  • Connections:
    • Center Computer Connector: RJ-45, 10M/100M, TCP/IP
    • Serial Port: RS232 serial port
    • Nurse Call: Nurse call connector
    • Equipotential Grounding Point: 1
    • USB Connector: Barcode scanner or other USB devices supported by the monitor
    • SD Card Connector: Reserved connector
    • Wireless Network: Reserved connector
    • Temp Module Connector: Temp probe connect
  • Data Storage:
    • Patient Quantity: 1000
    • Recording Number: 16000
    • Trend Data: 3 kinds resolution (1Min, 5Min, 10Min)
      • 1 Min: 96 hours
      • 5 Min: 480 hours
      • 10 Min: 960 hours
    • Alarm events: 1000
    • NIBP Measurement Record: 5000
  • SpO2:
    • Type: Digital, Nellcor, Masimo
    • Range: 0 to 100%
    • Digital PR Range: 25 ~ 255 bpm
    • Nellcor PR Range: 20 ~ 300 bpm
    • Masimo PR Range: 25 ~ 240 bpm
  • NIBP:
    • Type: Automatic oscillometry
    • Cuff Pressure Range: 0 ~ 300 mmHg
    • PR Range: 40 ~ 240 bpm
    • PR Resolution: 1 bpm
    • Measurement Time: 20 ~ 45 seconds
  • CO2:
    • Type: Infrared microstream
    • Range: 0 ~ 19.7% (0 ~ 150 mmHg)
    • Resolution: 0.1% or 1 mmHg
  • TEMP:
    • Ambient Temperature: 10? ~ 40?
    • Measurement Range: 34.0?~43.0?
    • Measurement Part: Ear Cavity
    • Unit: ?/?
    • Resolution: 0.1?/?
    • Accuracy: Between 35.0?~42.0?: =±0.2?; Beyond 35.0?~42.0?: =±0.3?