Description
DESCRIPTION: Bovie IDS 400 Features
- The Bovie IDS 400 integrates IDS technology, offering surgeons flexibility, convenience, and safety in choosing the desired effect consistently.
- Innovative electrosurgical generator with fully digital implementation for use in modern OR and surgical outpatient centers.
- Capable of covering a spectrum of electrosurgical interventions with monopolar and bipolar functions, ensuring maximum safety and user-friendly interface.
- BovieFDFS (Fast Digital Feedback System): Digitally controlled CUT and COAG functions provide consistent clinical output.
- BovieNEM (Neutral Electrode Monitoring).
- BovieFCFS (First Come First Serve): Improved patient safety with the latest digital implementation.
- BovieDED (Digital Error Detection).
SPECIFICATIONS: Bovie IDS 400 Specifications
- Dimensions:
- Height: 6.5 In (16.5 cm)
- Width: 14.75 In (37.5 cm)
- Depth: 18.1 In (46 cm)
- Weight: < 20 lbs (< 9.07 kg)
- General:
- Duty Cycle: 10 S / 30 S
- Output Type: RF Isolated “F”
- Output Frequency: 490 kHz ± 5
- Input Current: 4.5 A ~
- Line Voltage: 100-240 V ~
- Operating Parameters:
- Ambient temperature range: 10° to 40° C
- Relative humidity: 30% to 75%, non-condensing
- Atmospheric pressure: 70kPa to 106kPa
- Warm-up time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.
- Transport and Storage:
- Ambient temperature range: -40° to +70° C
- Relative humidity: 10% to 100%, including condensation
- Atmospheric pressure: 50kPa to 106kPa
- Modes:
- Cut I:
- Output Power: 400 W @ 500Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Crest Factor at Rated Load: 1.6 ± 20%
- VPeak Max: 1000V
- Cut II:
- Output Power: 300 W @ 300Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Crest Factor at Rated Load: 1.6 ± 20%
- VPeak Max: 625V
- Blend (Max):
- Output Power: 200 W @ 300Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Repetition Rate: 30 kHz ± 5 kHz
- Crest Factor at Rated Load: 3.5 ± 20%
- VPeak Max: 2000V
- Pinpoint:
- Output Power: 120 W @ 500Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Repetition Rate: 30 kHz ± 5 kHz
- Crest Factor at Rated Load: 4.5 ± 20%
- VPeak Max: 2400V
- Spray:
- Output Power: 80 W @ 500Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Repetition Rate: 30 kHz ± 5 kHz
- Crest Factor at Rated Load: 6.5 ± 20%
- VPeak Max: 4000V
- Bipolar:
- Output Power: 120 W @ 150Ω
- Output Frequency: 490 kHz ± 4.9 kHz
- Crest Factor at Rated Load: 1.6 ± 20%
- VPeak Max: 450V
- Cut I:
Top of Form