Description
- Description:
- The CareFusion Alaris BD 8110 Pump and Syringe module is designed for fluid delivery to patients, supporting intravenous, intra-arterial, and enteral routes.
- It utilizes a pressure-sensing dispenser for accurate medication delivery, ensuring precise dosing and infusion control.
- Compatible with the Alaris PC unit and Guardrails drug library system, enabling seamless integration and medication safety protocols.
- Lightweight design with front-loading syringe mechanism for easy operation and handling.
- Durable construction capable of supporting up to four additional modules for expanded functionality.
- Specifications:
- Bolus Volume, Maximum after Occlusion:
- Without Pressure Sensing Disc:
- Pressure Setting Low: 0.512 Bolus Volume (mL)
- Pressure Setting Medium: 0.776 Bolus Volume (mL)
- Pressure Setting High: 1.103 Bolus Volume (mL)
- With Pressure Sensing Disc:
- Pressure Setting 300 mmHg: Back Off Disabled: 0.462; Back Off Enabled: 0.126
- Pressure Setting 500 mmHg: Back Off Disabled: 0.575; Back Off Enabled: 0.331
- Pressure Setting 1000 mmHg: Back Off Disabled: 0.839; Back Off Enabled: 0.323
- Critical Volume: Maximum over-infusion during a single-fault condition is within 2% of nominal syringe fill volume during loading.
- Flow Rate Range: 0.01-999 ml/hr
- Dimensions:
- Width: 4.5″
- Height: 15″
- Depth: 7.5″
- Weight: 4.5 Lbs.
- Without Pressure Sensing Disc:
- Bolus Volume, Maximum after Occlusion: