Description
ConMed Excalibur Plus Features:
- Most Powerful Isolated ESU: The ConMed Excalibur Plus PC is the most potent isolated Electrosurgical Unit (ESU) available in its class, providing exceptional performance and versatility in surgical procedures.
- Adjustable Power Settings: Surgeons can conveniently adjust power settings from the surgical field using a standard ES pencil, enhancing control and efficiency during procedures.
- Patented Controlled Output Voltage in CUT: The Excalibur Plus PC features patented technology that maintains the minimum necessary voltage for smooth cutting through various impedance tissues, minimizing thermal damage and ensuring precision.
- Aspen Return Monitor (ARMU): Incorporates the Aspen Return Monitor (ARMU), the most sensitive pad monitoring system available, which combines automatic feedback circuitry and visual displays to provide precise contact quality monitoring during surgeries.
- Electrosurgical Generator with Multiple Modes: Offers Cut, Coag, Bipolar, and Blend modes, providing flexibility and adaptability for various surgical applications.
- Output Power: 300 watts of output power, ensuring sufficient energy delivery for demanding procedures.
- Scalpel-like Cut: The Excalibur Plus PC delivers a scalpel-like cut and offers three blend modes for enhanced surgical precision and control.
- Peak Voltage Specifications:
- Pure Cut: Maximum Peak to Peak Voltage of 2000 Volts
- Blend 1: Maximum Peak to Peak Voltage of 2000 Volts
- Blend 2: Peak Voltage of 2200 Volts
- Blend 3: Maximum Peak to Peak Voltage of 2400 Volts
SPECIFICATIONS: ConMed Excalibur Plus
- Dimensions:
- Height: 6.25 in (15.9 cm)
- Width: 18 in (45.8 cm)
- Length: 22.75 in (57.8 cm)
- Weight: 40 lbs. (18.1 kg)
- Environmental Conditions:
- Operating Conditions: 10°C to 30°C, 95% RH non-condensing
- Storage Conditions: -14°C to 70°C, altitude -60 to +4500 m MSL when sealed in original poly bag
- Line Regulation: Power change < 1% /V for vline = min to max voltage
- Power Display Accuracy: The greater of +/- 10% of display or +/- 3 watts rated load.
- Pad Monitor:
- Single Pad Mode: Two-wire continuity detector, typical threshold = 10 Ohms
- Dual Pad Mode: Two-wire resistance monitor, typical acceptance range 10 to 150 OHMS, trip threshold typically 20% higher than the last monitor set point activation, visual indication of patient resistance.
- Activation Frequencies:
- Cut: 520 HZ
- COAG: 400 HZ
- Operator Error: 1710 HZ, pulsating
- Plate Alarm: 1000 HZ (65 DBA MINIMUM)
- Machine Fault: 1000 HZ (45 DBA Min. Except Alarms Are 65 DBA Min)