Description
Conmed Hyfrecator 2000 Features:
- Wide Range of Applications: Suitable for dermatology, gynecology, ophthalmology, urology, and various other medical procedures.
- Dual Microprocessor Technology: Utilizes two microprocessors to ensure precise output power and accuracy while prioritizing patient safety.
- Compatibility: Compatible with existing Hyfrecator accessories such as bipolar forceps and reusable electrodes, providing versatility and cost-effectiveness.
- State-of-the-Art Electronic Circuitry: Incorporates advanced electronic circuitry to deliver accurate output power, offering flexibility from broad surface coagulation to delicate facial procedures.
- Internal Self-Diagnostic Tests: Performs internal self-diagnostic tests upon powering on to ensure proper functionality and safety.
- Ergonomic Design: Designed with user comfort and convenience in mind, facilitating ease of use during medical procedures.
- Precise Power Delivery: Offers precise and accurate power delivery, with the ability to adjust power settings in 1/10-watt increments for delicate procedures.
- Lightweight: Weighs only 6 lbs (2.7 kg), making it portable and easy to transport between medical settings.
SPECIFICATIONS: Conmed Hyfrecator 2000
- Dimensions:
- Height: 7.5” (19 cm)
- Width: 8.75” (22.2 cm)
- Depth: 4” (10.2 cm)
- Weight: 6 lbs. (2.7 kg)
- Power Accuracy at Rated Load:
- LO and BI: 0 to 10 W: Greater of 90 mW or 10% of setting; 10 to maximum settings: 10% of setting
- HI: Greater of 1W or 10% of setting
- RF Frequency: 450 kHz ±50kHz (open circuit)
- Output Wave Shape: Clipped, damped sinusoidal
- Line Regulation: ±5% at max. power for a ±10% line voltage
- Duty Cycle: Intermittent: 30 sec. on, 30 sec. off
- Audio: Adjustable from >45 dba to =65 dba at 1 meter
- 60 Hz Leakage: 50 microamps max. third wire of the line cord open
- Display: Lower output Terminal; 0-10 Range: 0.1 watt increments; 10-20 Range: 1-watt increments; Bipolar & High Output Terminals: 0-35 Range: 1-watt increments
- Functional Characteristics: Hi Mode, LO Mode, BI Mode with respective specifications
- Environmental Conditions:
- Operating Temperature Range: +10 degrees to +40 degrees C
- Relative Humidity: 10% to 95% non-condensing
- Storage Restrictions: -40 degrees to +70 degrees C, 95% humidity N.C
- Operating Conditions: 10°C to 30°C, 95% RH Non- non-condensing maximum at altitudes between 60 meters below and 3000 meters above mean sea level