Conmed Sabre 2400

Single Product Request Form
Category

Description

Conmed Sabre 2400 Features:

  • Full-Power Performance: The ConMed Sabre 2400 is a full-power electrosurgical generator designed to accommodate the needs of all surgical procedures, meeting the highly demanding performance requirements of electrosurgery.
  • Independent Outputs: Features independent monopolar outputs and enhanced bipolar performance, providing versatility and efficiency during surgical procedures.
  • Thermal Damage-Free Cutting: Incorporates patented controlled output voltage technology in the cut mode, ensuring superior cutting performance without causing thermal damage to tissues. The voltage is constantly adjusted as the electrosurgical electrode encounters different tissue types, delivering precise and efficient cutting.
  • Blend Mode: Surgeons can utilize the Blend mode to achieve greater hemostasis while cutting, enhancing control and effectiveness during procedures.
  • Auto Spray Feature: Offers superior spray coagulation at power settings above 50 watts, providing efficient and controlled coagulation.
  • Enhanced Bipolar Mode: Provides both hand and foot-switching bipolar coagulation with minimal tissue sticking, ensuring optimal performance and outcomes.
  • High-Power Performance: Delivers high-power performance and key user features to meet demanding electrosurgical needs, designed to meet the requirements of even the most demanding surgeons in various surgical applications.
  • Safe and Cost-Effective: Safe, convenient, and cost-effective, the Sabre 2400 can be programmed for standardization, contributing to efficient and standardized surgical practices.

SPECIFICATIONS: Conmed Sabre 2400

  • Dimensions:
    • Height: 6.75″ (17.1 cm)
    • Width: 14.25″ (36.2 cm)
    • Depth: 21.5″ (54.6 cm)
    • Weight: 32 lbs (14.5 kg)
  • General:
    • Pure Cut:
      • Max Power: 300 W
      • P-P Voltage: 2000
      • Load (ohms): 300
      • Crest Factor: 1.8
    • Blend Cut:
      • Max Power: 180 W
      • P-P Voltage: 2200
      • Load (ohms): 300
      • Crest Factor: 2.5
    • Coag:
      • Max Power: 120 W
      • P-P Voltage: 9000
      • Load (ohms): 500
      • Crest Factor: 7.0
    • Bipolar:
      • Max Power: 50 W
      • P-P Voltage: 280
      • Load (ohms): 50
      • Crest Factor: 1.5 – 12.1
    • Dispersive Electrode Monitor:
      • Aspen Return Monitor alarms above 150 ohms, below 10 ohms, and with a 20% increase in resistance above the set point.
    • Operating Frequency:
      • Cut: 416.7 kHz ± 0.01%
      • Coag: 416.7 kHz ± 0.01%
    • Pulse Repetition Frequency:
      • Blend: 20.0 kHz ± 0.01%
      • Coag: 25.0 kHz ± 0.01%
      • Bipolar: 20.0 kHz ± 0.01%
    • Input Voltage:
      • 104-127 VRMS, 50/60 Hz, calibrated at 120V
    • Input Current:
      • Pure cut at 300: 7.0A max
    • Line Frequency Leakage:
      • Patient connections to neutral: < 25 μA
      • Chassis to neutral: < 25 μA typical
    • RF Leakage Current:
      • Less than 100 mA (200 ohms to ground)
    • Power Decrease:
      • 1% typical. Pure cut set to 150 watts for 20 seconds