Description
- Specifically designed for hospital use, surgery monitoring, bedside, or transport monitoring.
- Large color display for easy reading, capable of displaying up to three waveforms simultaneously.
- Measures parameters including EKG, NIBP, pulse oximetry, temperature, and can be configured for end-tidal CO2 monitoring.
- Features 3/5 Lead ECG, printer, non-invasive blood pressure, and battery backup for transport.
Datascope Passport XG Specifications:
- Display:
- Size: 21.1 cm x 15.8 cm
- ECG:
- Leads: I, II, III, aVR, aVL, aVF, and V.
- Defibrillator Overload Protection: Withstands Lown and trapezoidal or Edmark waveform up to 400 joules applied to 100 ohms load and electrode at 20 second intervals.
- Recovery Time: 3 seconds maximum.
- IBP:
- Range: Systolic/Diastolic Mean -20 to +300 mmHg.
- Accuracy: ±2 mmHg or 2%, whichever is greater.
- Heart Rate:
- Range: 30-300 bpm for adult/child, 30-350 bpm for neonatal.
- Accuracy: ±3 bpm or ±3%, whichever is greater.
- NIBP:
- Technique: Oscillometric.
- Systolic Range: 50 to 250 mmHg for adult/ped mode, 15 to 150 mmHg for neonatal mode.
- Diastolic Range: 30 to 200 mmHg for adult/ped mode, 10 to 150 mmHg for neonatal mode.
- Accuracy: Mean error less than 7 mmHg, standard deviation less than 8 mmHg (adult/ped), mean error less than 4 mmHg, standard deviation less than 13 mmHg (neonatal).
- Temperature:
- Scale: Selectable °C or °F.
- Range: 15°C to 45°C, 59°F to 113°F.
- Accuracy: ±0.1°C to 45°C or ±0.2°F 59 to 113°F.
- Respiration (ECG):
- Range: 4-200 bpm.
- Accuracy: ±2% or ± bpm, whichever is greater.
- Recorder:
- Speed: 50, 25, 12., 6.25, 3.125 mm/sec ±5%.
- CO2:
- Range: 0-14.8 vol%, 0-14.8 kPa, 0-11 mmHg, 0-111 Torr.
- Rise Time: <400ms.
- Accuracy: ±(0.2% vol% to 2% of reading).
- O2:
- Range: 0-100 vol%.
- Rise Time: <400 ms.
- Accuracy: ±(0.2% vol% + 2% of reading).
- N2O:
- Range: 0-100% N2O.
- Rise Time: <450 ms.
- Accuracy: ±(2 vol%+ 2% of reading).
- Battery:
- Type: 2 sealed lead acid batteries.
- Voltage: 12 VDC nominal (@2.3A load).
- Capacity: 2.3 AH each.
- Run Time: 1.5 hours from full charge of new battery at 25°C with specified monitoring parameters.
- Recharge Time: 16 hours.
- Low Battery Warning: 11.6V, ±3.5%.
- Low Battery Cut Off: 11.0V ±2.8%.