Description
Features:
- Designed for use in operating rooms, the Datascope Spectrum OR is a patient monitor providing essential measurements for surgical environments.
- Offers readings for ECG, SpO2, NIBP, Spirometry, temperature, and BISx for comprehensive patient monitoring.
- Based on the same platform as the Passport 2, it offers either Masimo or Nellcor SpO2 monitoring.
- Features a 12-inch high-contrast color display for optimal visibility with quick-access menus and one-touch functionality.
- Includes pre-anesthesia monitoring features like 3 or 5 lead ECG, non-invasive blood pressure, and either Nellcor or Masimo SpO2 monitoring.
Datascope Spectrum OR Specifications:
- Dimensions:
- Height with Module: 10.7 in (27.2 cm)
- Width with Module: 11.9 in (30.2 cm)
- Depth with Module: 7.5 in (18.8 cm)
- Weight:
- With Modules: 1.75 lb (0.8 kg)
- With Battery: 13.52 lb (6.1 kg)
- Display:
- Size: 12.1-inch (30.7 cm) Color Active Matrix TFT LCD
- Resolution: 800 x 600 pixels
- Waveforms: 3 to 8
- ECG:
- Leads: I, II, III, aVR, aVL, aVF, V
- Display Sensitivity: 0.25, 0.5, 1.0, 2.0, 3.0, 4.0 cm/mV ± 10%
- Heart Rate Meter:
- Range: 30 – 300 BPM Adult/Pediatric, 30 – 350 BPM Neonate
- Accuracy: ± 3 BPM or ± 3% at 30 – 250 BPM whichever is greater, ± 5% at 251 – 350 BPM
- Respiration (ECG):
- Range: 4 to 199 BPM
- Accuracy: ± 2% or 2 breaths per minute whichever is greater from 4 to 150 BPM, ± 4% from 151 to 199 BPM
- Non-Invasive Blood Pressure (NIBP):
- Systolic Range: Adult: 55 – 235 mmHg, Pediatric: 55 – 160 mmHg, Neonate: 45 – 120 mmHg
- Diastolic Range: Adult: 30 – 200 mmHg, Pediatric: 30 – 150 mmHg, Neonate: 20 – 100 mmHg
- Accuracy: Mean Error less than ± 5 mmHg, standard deviation less than ± 8 mmHg
- Pulse Oximetry:
- Masimo SET SpO2 Accuracy Saturation with no motion conditions: Adult/Pediatric: 70% to 100% ± 2 digits SpO2
- Neonate: 70% to 100% ± 3 digits SpO2
- CO2 (Microstream):
- Range: 0 – 99 mmHg
- Accuracy: 0 – 38 mmHg ± 4 mmHg, 39 – 99 mmHg ± 12%
- Battery:
- Type: Sealed Lead Acid
- Voltage: 12 V
- Capacity: 2.3 Ah
- Run Time: 1 hour, 20 min from two fully charged, new batteries at 25°C with specified monitoring parameters