Features:

  • Full antepartum/intrapartum fetal monitoring system.
  • Monitors fetal heart rate (FHR) and uterine activity (UA), along with maternal parameters such as blood pressure, temperature, oxygen saturation, ECG, and non-invasive blood pressure.
  • Equipped with a color LCD display for simultaneous display of fetal and maternal parameters, including maternal waveforms.
  • Recorder prints continuous trends and alphanumeric data on one strip chart.
  • Compatible with Centricity Perinatal Clinical Information Systems and other information systems for data capture and archiving.
  • Supports two FHR channels, uterine activity (TOCO or IUP), maternal non-invasive blood pressure (NIBP), maternal pulse oximetry (MSpO2), fetal ECG (FECG), and maternal ECG (MECG).
  • Backlit keys and dimmable display for ease of use.
  • Utilizes 9-crystal ultrasound technology for accurate fetal monitoring.
  • One-touch hot keys for quick access to essential functions.

Specifications:

  • Dimensions:
    • Height: 6.7″ (17 cm)
    • Width: 16.7″ (42.4 cm)
    • Depth: 17.5″ (44.4 cm)
    • Weight: 22 lbs (10.9 kg)
  • Power:
    • Nominal Line Voltage: 100 VAC, 120 VAC, 220 VAC, 230 VAC, 240VAC
    • Line Frequency: 50/60Hz
    • Power Consumption: 100 W
    • Chassis Leakage: <300 µA
  • Environmental:
    • Operating:
      • Ambient Temperature: 50°F to 104°F (10°C to 40°C)
      • Relative Humidity: 10% to 95%, non-condensing
      • Atmospheric Pressure: 700–1060 mbar (525–795 mmHg)
    • Storage:
      • Ambient Temperature: 14°F to 131°F (–10°C to 55°C)
      • Relative Humidity: 0% to 95%, non-condensing
      • Atmospheric Pressure: 700–1060 mbar (525–795 mmHg)
    • Recorder Drive:
      • Speed: 1, 2, and 3 cm/min
      • Speed Accuracy: ± 1%
    • FECG:
      • Technique: Peak detecting, beat-to-beat cardio tachometer
      • Heart Rate Counting Rage: 30-240 bpm
      • Heart Rate Resolution: ± 1 bpm
    • Ultrasound Mode:
      • Technique: Pulsed Doppler with autocorrelation processing
      • Transducer Type: 9-crystal
      • Pulse Repetition Frequency:
        • Single Ultrasound Mode: 4 kHz
        • Dual Ultrasound Mode: 2 kHz
      • Uterine Activity Mode:
        • Toco transducer
        • Range (typical): 0–100 mmHg (0–13.3 kPa)
        • Resolution: 1 mmHg (0.13 kPa)
      • MECG:
        • Technique: Peak detecting, beat-to-beat cardiotachometer
        • Leads Available: I, II, and III
        • Heart Rate Counting Range: 30–240 bpm
        • Heart Rate Resolution: ± 1 bpm

Maternal Blood Pressure Mode:

  • Technique: Oscillometric w/DINAMAP; SuperSTAT algorithm
  • Blood Pressure Range: Systolic 30-290 mmHg, Diastolic 10-220 mmHg, Mean arterial pressure 20-260 mmHg
  • Pulse Rate Range: 30-200 BPM
  • Blood Pressure Accuracy: ±5 mmHg with a standard deviation no greater than 8 mmHg
  • Pulse Rate Accuracy: ±2 BPM or ±2% (whichever is greater)