Philips Avalon CL

Single Product Request Form
Category

Description

Philips Avalon CL Features:

  • Wireless fetal monitor system for continuous monitoring during antepartum, labor, and delivery.
  • Cableless transducers allow for easy movement of the patient without sacrificing monitoring accuracy.
  • Reduction of stress caused by wires for both mother and baby.
  • Transducers transmit fetal and maternal vital signs wirelessly to a monitor at a base station in the medical facility.
  • Measures fetal heart rate of twins and triplets, along with maternal NIBP and SpO2.
  • Smart Pulse technology enhances safety measures for maternal parameters when using the Toco transducer.
  • Uses CCV to compare heart rates of mother and baby to reduce the risk of mistaking one heart rate for another in cases of multiple heart rates.
  • Allows monitoring of multiple heart rates, twins, triplets, and the mother.
  • Cabless, waterproof, and lightweight transducers.
  • Enables continuous monitoring during bathing or ambulation.
  • Provides on-demand audible patient calls from the fetal monitor.
  • Reliable fetal parameter signal quality with a line-of-sight operating range of up to 100 m/300 ft.
  • Automatic out-of-range notification for the patient.
  • User-friendly intuitive controls at the display of the connected fetal monitor.
  • High-capacity, rechargeable Lithium-Ion batteries offer more than 8 hours of operating time per charge cycle with a recharge time of less than 3 hours.
  • Easy and quick assignment of transducers and CL Pods by simply docking them at the base station.
  • The base station connects to any fetal sensor socket on the Philips Avalon FM20, FM30, FM40, and FM50 fetal monitors or to the rear telemetry connectors on the FM40 and FM50.

Philips Avalon CL Specifications:

  • Dimensions:
    • Height: 2.9 in (74 mm)
    • Width: 13.7 in (349 mm)
    • Depth: 7.2 in (183 mm)
    • Weight: 2.2 lb (0.985 kg)
  • Transducer Dimensions:
    • Toco + MP:
      • Size (diameter/height): 2.9/1.4 in (75.5/36.6 mm)
      • Weight: 123 g ±10% (4.3 oz ±10%)
    • US Transducer:
      • Size (diameter/height): 2.9/1.4 in (75.5/36.6 mm)
      • Weight: 131 g ±10% (4.6 oz ±10%)
    • Transducer Battery:
      • Type: Integrated rechargeable li-ion battery with a battery gauge and cycle counter
      • Runtime (fully charged battery): Minimum eight hours
      • Life: > 500 charges/discharge cycles or older than 4 years
      • Transducer Storage Time: ≤ 1 year at 25°C (77°F) (battery should be charged to 40% – 50% every six months)
      • Recharge Time: 100% charge = 3.0h
      • Recharging Time with an FM20/30: 100% charged > 6.0h
    • Environmental (Transducers and Base Station):
      • Operating temperature range: 0-45°C (32-113°F)
      • Operating transducer temperature: 0-40°C (32-113°F)
      • Storage transducer temperature: -20°C- +60°C (-4-140°F)
      • Operating humidity range: <95% relative humidity @ 45°C (113°F)
      • Storage humidity range: <95% relative humidity @ 60°C (140°F)
    • CL:
      • Intensity:
        • Average output power: P = (12.4 ± 0.4) mW
        • Peak-negative acoustic pressure: P_ = (49.1 ± 5.2) kPa
        • Output beam intensity(= spatial average – temporal average intensity): Isata = (2.77 ± 0.56) mW/cm²
        • Spatial-peak temporal average intensity: Isata=(21.1 ± 5.1) mW/cm²
      • Effective radiating area @ -12 dB: A-12 dB = (4.47 ± 0.89) cm²
      • Beat-to-beat change (max. for Ultrasound): 28 bpm
      • US frequency: 1 MHz ± 100Hz
      • US signal range: 3.5µ Vpp to 350 µ Vpp @ 200 Hz
      • US burst repetition rate: 3.0 kHz
      • US burst duration: ≤ 100 µs
    • Toco:
      • Measurement Method: Strain Gauge Sensor Element
      • Sensitivity: 1 unit = 2.5 g
      • Measurement Range: 400 units
      • Signal Range: 0-127 units
      • Maximum offset range: -300 units
      • Baseline setting: 20 units
      • Auto Offset Connection: 3 seconds after connecting the transducer, the Toco value is set to 20 minutes
      • Auto Zero Adjust: Toco value is set to zero following a negative measurement value for 5 seconds
    • Maternal Pulse from Toco:
      • Emitted light energy: ≤ 15m W
      • Wavelength range: 780-1100nm
      • Range:40-240 bpm
      • Accuracy: ±2% or 1bpm, whichever is greater
      • Update rate: every 4 seconds