Description
Philips Avalon CTS M1351a Features:
- Cordless transducer system for fetal and maternal monitoring, providing flexibility and mobility during labor and delivery.
- Includes cordless toco transducer for measuring uterine activity externally.
- Cordless ultrasound transducer measures fetal movement and heart rate.
- Bed identifier matches transducers and base stations for each bed, ensuring accurate monitoring.
- Base station receives transducer signals and recharges plugged-in transducers.
- Transducer plug and play with automatic screen layout simplifies operation and allows focus on patient care.
- Transducer status indicators provide real-time feedback on system performance.
- Slot position indicator ensures each transducer is returned to its proper home position.
- Cordless ECG transducer measures direct fetal ECG or maternal ECG.
- Battery charge indicator shows the status of charge, with batteries operating for at least 16 hours.
Philips Avalon CTS M1351a Specifications:
- Base Dimensions:
- Width: 350 mm (13.8 in)
- Depth: 240 mm (9.5 in)
- Height: 75 mm (3 in)
- Weight: 2.5 kg/5.5 lbs without transducers
- Transducer Dimensions:
- Size (diameter): < 10 cm/3.94 in
- Weight: < 140 g/4.8 oz
- Power:
- Supply Voltages: 100 VAC to 240 VAC ± 10%
- Supply Frequency Range: 50 Hz to 60 Hz
- Consumption: 15 VA
- Base Station:
- Type of Protection Against Electrical Shock: Class I equipment
- Input Sensitivity: -110 dBm @ 30dB Signal-to-Noise Ratio
- Image Rejection: > 80 dB
- Receiving Range (line of sight): approximately 100 m/300 ft
- Antenna Input Impedance: 50 Ω
- Water Ingress Protection Code: IP X1 (protection against vertically falling water drops)
- Monitor Interface:
- Toco Output Accuracy: ± 0.5% per 100 mmHg
- Toco Output Range: 0 to 4V
- US Voltage Range: 4mVpp to 4Vpp
- ECG Voltage Range: 0.1Vpp to 4Vpp
- Battery:
- Type: Lithium Ion
- Capacity: > 16 hours (> 700mAh)
- Life: > 500 charge/discharge cycles
- Transducer Storage Time: ≤ 1 year at 25°C/77°F (battery full), ≤ 1 month at 25°C/77°F (battery empty)
- Recharging Time: 100% charged < 3h, 66% charged = 1h
- RF Unit:
- Nominal RF Output Power: -12 dBm ± 6 dB ERP
- Carrier Frequency Range: See Frequency Bands
- Minimum Frequency Band Span Per Option: 10 MHz
- Channel spacing: 25 kHz or 12.5 kHz
- General Transducer Specifications:
- Shock Resistance: Withstands a 1m drop to concrete surface
- Usability Underwater: 0.5m/1.64 ft
- Water Ingress Protection Code: IP 68 (0.5m/1.64 ft immersion for 5 hours)
- Environmental:
- Temperature Range:
- Charging: 0°C to 45°C (32°F to 113°F)
- Operating: 0°C to 45°C (32°F to 113°F)
- Storage (without battery): -20°C to +60°C (-4°F to 158°F)
- Humidity Range:
- Operating: 5% to 95% relative humidity @ 40°C/104°F
- Storage: 5% to 85% relative humidity @ 50°C/122°F
- Altitude Range:
- Operating: ≤ 3000m/9800 ft
- Storage: ≤ 15000m/49000 ft
- Temperature Range: