Philips Avalon FM30

Single Product Request Form
Category

Description

Philips Avalon FM30 M2703a Features:

  • Fetal and maternal monitor offering cross channel verification to separate maternal pulse rate and blood pressure from fetal heart rate.
  • Color touchscreen interface displaying ultrasound, fetal movement, direct fetal heart rate, TOCO, and intrauterine pressure, as well as maternal pulse rate, ECG, SpO2, and blood pressure.
  • Cordless transducer system allowing freedom of movement for the mother during continuous monitoring.
  • Intuitive color touchscreen with large numerics and graphics for easy interpretation.
  • Backup memory for seamless monitoring.
  • LAN interface for compatibility with hospital IT networks and access to additional system capabilities.
  • Smart transducers with auto-recognition simplify operation.
  • Transducer plug and play with automatic screen layout to focus on the patient.
  • Data buffer with multi-patient handling for random printout.
  • Cross-channel verification of maternal and fetal heart rates.

Philips Avalon FM30 M2703a Specifications:

  • Dimensions (M2702A/M2703A):
    • Height: 5.2 in (133 mm)
    • Width: 11.3 in (286 mm)
    • Depth: 13.2 in in (335 mm)
    • Weight: 11.2 lbs (5.1 kg)
  • Dimensions (M2704A/M2705A):
    • Height: 6.8 in (172 mm)
    • Width: 16.5 in (420 mm)
    • Depth: 14.6 in (370 mm)
    • Weight: 19.8 lbs (<9.0 kg)
  • Display:
    • Monitor Screen: 6.5 in / 16.51 cm
    • Touchscreen Operation: Optional
  • Battery:
    • Operating Time: up to 4 hours
  • Environmental Conditions (Monitor Operating):
    • Temperature Range: 0°C to 45°C (32°F to 113°F)
    • Humidity Range: <95% relative humidity @ 40°C/104°F
    • Altitude Range: -500 to 3000 m / -1640 to 9840 ft.
  • Environmental Conditions (Monitor Storage):
    • Temperature Range: -20°C to 60°C (-4°F to 140°F)
    • Humidity Range: <90% relative humidity @ 60°C/140°F
    • Altitude Range: -500 to 13100 m / -1640 to 43000 ft.
  • Ultrasound (Fetal / Maternal):
    • Measurement Method: Ultrasound Pulsed Doppler
    • Measurement Range: 50 to 240 bpm
    • Display Resolution: 1 bpm
  • ECG:
    • Measurement Range: 30 to 240 bpm
    • Display Resolution: 1 bpm
  • Non-invasive Blood Pressure:
    • Measurement Ranges: Systolic 30 to 270 mmHg, Diastolic 10 to 245 mmHg, Mean 20 to 255 mmHg
    • Accuracy: Max. Std. Deviation: 8 mmHg, Max. Mean Error: ±5 mmHg
  • Pulse Rate Range Accuracy:
    • 40 to 300 bpm: 40 to 100 bpm: ±5 bpm, 101 to 200 bpm: ±5% of reading, 201 to 300 bpm: ±10% of reading
  • Transducers and Measurements: Includes ultrasound fetal heart rate monitoring, fetal movement profile detection, external uterine activity monitoring, Smart Pulse technology for cross-channel verification, NST timer for antepartum applications, and optional maternal noninvasive blood pressure monitoring.
  • Monitor: Features high-resolution color TFT touchscreen display, intuitive user functionality via touchscreen, automatic screen layouts, and optional battery operation.
  • General Information About Fetal Monitors: Provides an overview of fetal monitoring terminology, monitoring modes (antepartum and intrapartum), and parameters to be considered.