Description
Features:
- Versatile vital signs monitoring with the flexibility to choose between NBP interval mode and spot check mode.
- Intuitive touch-screen interface with tools like QuickCapture and QuickCheck for simplifying clinician workflow.
- Wireless capability and proactive monitoring tool QuickAlert support Rapid Response teams.
- Compatible with existing networks for easy integration.
- Touch screen with color LED backlight.
- Internal WiFi/802.11 a/b/g option.
- Programmable 2D barcode scanner for multiple ID fields.
- Time-sync with hospital or network clock.
- Improved battery management with visible charging indicator.
- Clearly identifiable one-touch NBP on/off button.
- Stores up to 800 patient records.
Specifications:
- Dimensions:
- Size: W: 26 cm (10.2”), H: 22 cm (8.6”), D: 14.5 cm (5.7”)
- Weight: 3.1 kg (6.9 lb) – excluding optional recorder
- Screen:
- 4” SVGA TFT-AM LCD display
- Resolution: 800 active pixels/line, 600 active lines per frame
- Viewing angle: ±60 degrees
- NBP (Non-invasive Blood Pressure):
- Oscillometric using stepwise deflation pressure
- Adult Measurement Range:
- Systolic: 30–270 mmHg (4.0–36.0 kPa)
- Diastolic: 10 –245 mmHg (1.3–32.7 kPa)
- MAP: 20–225 mmHg (2.7–34.0 kPa)
- Pediatric Measurement Range:
- Systolic: 30–180 mmHg (4.0–24.0 kPa)
- Diastolic: 10–150 mmHg (1.3–20.0 kPa)
- MAP: 20–160 mmHg (2.7–21.3 kPa)
- Neonatal Measurement Range:
- Systolic: 30–130 mmHg (4.0–17.0 kPa)
- Diastolic: 10–100 mmHg (1.3–13.3 kPa)
- MAP: 20–120 mmHg (2.7–16.0 kPa)
- Accuracy:
- Maximum Standard Deviation: 8 mmHg
- Maximum Mean Error: ±5 mmHg
- SpO2:
- Measurement Range: 0–100%
- Pulse Rate Measurement Range: 30–300 bpm
- Accuracy: Depends on Sensor
- Temperature:
- Welch Allyn Predictive:
- Modes: Predictive and Monitored
- Probe Sites: Oral, Rectal, or Axillary
- Range: 26.7–43.3°C (80–110°F)
- Accuracy: ±0.1°C (±0.2°F) in monitoring mode
- Covidien Tympanic:
- Probe Sites: Oral, Rectal, or Axillary
- Range: 33°C–42.0°C (91.4°F–107.6°F)
- Accuracy: ±0.2°C (36.7–38.9)
- Measurement response time: ≤2 seconds
- Battery:
- Lithium-ion: 10.8–11.1V (with a “202” form factor)
- Operating time: 4 hours with NBP every 15 minutes.
- Data Output:
- HL7 format, via Ethernet port
- Serial data
- Patient Type:
- Adult, Pediatric, Neonatal
- Welch Allyn Predictive: