Stryker Physio-Control Lifepak CR Plus

Single Product Request Form
Category

Description

Stryker Physio-Control Lifepak CR Plus Features:

  • Child Capability: Can be used on patients as small as 55 lbs or 8 years old with Child Reduced Energy Electrodes.
  • Biphasic Truncated Exponential Waveform: Utilizes ADAPTIV Biphasic technology for defibrillation.
  • Energy Range: Offers energy levels ranging from 150 to 360 joules.
  • Quick Charge Times: Charges to 200 joules in less than 9 seconds, and to 360 joules in less than 15 seconds.
  • Wireless Data Transfer:*Allows for data transfer to a computer wirelessly for easy analysis.
  • Lightweight: Weighs only 4.5 lbs (2 kg).

Stryker Lifepak CR Plus Specifications:

Dimensions:

  • Height: 4.2” (10.7 cm)
  • Width: 8” (20.3 cm)
  • Depth: 9.5” (24.1 cm)
  • Weight: 4.5 lbs (2 kg)

Defibrillator:

  • Waveform: Biphasic Truncated Exponential
  • Output Energy: 150 joules to 360 joules
  • Shock Advisory System: Provides advice on whether a shock is appropriate based on ECG analysis
  • Device Capacity: 30 full discharges or 210 minutes of “ON time”

Shock Charge Time:

  • 200 joules in less than 9 seconds
  • 360 joules in less than 15 seconds

CHARGE-PAK Battery Charger:

  • Type: Li/SO2Cl2 Lithium Sulfuryl Chloride, 11.7V, 1.4 amp-hours
  • Replacement: Replace after each use or when the CHARGE-PAK indicator is visible (typically after 2 years)
  • Weight: 80.5 grams (0.18 lb)

User Interface:

  • Includes voice prompts, audible tones, and graphic prompts
  • Readiness Display: Shows the device status
  • OK Indicator: Indicates successful completion of the last self-test
  • CHARGE-PAK Indicator: Indicates the need to replace the CHARGE-PAK battery charger
  • Attention Indicator:  Indicates remaining discharges or operating time
  • Service Indicator: Indicates when service is required

Default Settings:

  • Energy sequence set to 200J, 300J, 360J
  • Motion detection system set to on during analysis
  • Energy protocol set to increase energy only after a lower energy shock was unsuccessful

Data Storage:

  1. Internal digital memory
  2. Dual patient data storage with a minimum of 20 minutes of ECG stored for the current patient

Environmental:

  • Operating Temperature: 0° to +50°C (+32° to +122°F)
  • Long-term storage temperature: +15° to +35°C (59° to 95°F)
  • Short-term storage temperature: -40° to +70°C (-40° to 158°F) for a maximum of one week
  • Water Resistance: IPX4 “Splash proof” with electrodes connected, CHARGE-PAK installed