Zimmer ATS 1200

Single Product Request Form
Category

Description

Zimmer ATS 1200 Features:

  • Provides accurate pressure monitoring for patient safety and confidence.
  • Microprocessor controls ensure reliability with built-in safety features and self-diagnostic tests.
  • Suitable for single port, dual cuff surgical tourniquet system.
  • Offers rapid inflation and deflation for efficient procedures.
  • Equipped with cuff lockout feature to keep the cuff inflated even if power is turned off.
  • Dual cuffs allow for bilateral or Bier Block procedures with critical monitoring.
  • Includes visual and audio alarms for enhanced safety.
  • Utilizes positive locking connectors for leak-free inflation.
  • Offers color-coded cuffs in various sizes, including low-profile options.
  • Lightweight unit can be easily used on a table or IV pole.

Zimmer ATS 1200 Specifications:

  • Dimensions:
    • Height: 16.13 inches
    • Width: 8.75 inches
    • Depth: 7.1 inches (Including 2 inches clamp)
  • Weight: 19.8 lbs
  • General:
    • Line Voltage Range: 105-130 VAC, 50/60 Hz
    • Line Current: 485 mA RMS Max., 300 mA RMS nominal
    • Input Power: 58 W max, 36 W nominal
    • Power Cord Type: SJT, AWG 18, 13 ft. 2 in (4.01m)
    • Power Plug: Hospital grade, 3 prongs straight blade, 15A
    • Line Protection: 0.5A, 250V, 5 x 20 mm SLO-BLO fuse
  • Cuff Pressure:
    • Range: 50-475 mmHg
    • Pressure Accuracy: ± 5 mmHg (50-475 mmHg)
    • Pressure Regulation: ± 6mm of set point
    • Maximum Pressure: 475 mmHg (Nominal operation), 600 mmHg (internally limited)
  • Timer:
    • Time Alarm Set Ranges: 0-240 minutes; 1 minute increments
    • Timer Accuracy: 0.25% ± 1 second
  • Battery:
    • Battery Type: Rechargeable, sealed lead-acid, 12 VDC, 2.3 AH
    • Battery Discharge Time: 15 minutes (minimum with fully charged battery)