Bovie IDS 210

Single Product Request Form
Category

Description

DESCRIPTION: Bovie IDS 210 Features

  • The Bovie IDS 210 is a 200-watt high power, operating room generator, by Bovie.
  • Incorporates the power and safety of a premium OR generator at significant savings to the medical facility.
  • Delivers 12 monopolar and bipolar RF energy modalities for high surgical precision.
  • Engineered with patient safety in mind, featuring Bovie NEM return split pad contact quality monitoring, Bovie FDFS tissue sensing technology, self-test circuits, and specialized modalities for laparoscopic and delicate procedures.
  • Two Cut Modes
  • Four Levels of Blend
  • Three Coagulation Modes
  • Three Bipolar Modes
  • Fast Digital Feedback System
  • Three Front-Panel Accessory Connections and two rear-panel footswitch connections

SPECIFICATIONS: Bovie IDS 210 Specifications

  • Dimensions:
    • Height: 6.5 In (16.5 cm)
    • Width: 14.75 In (37.5 cm)
    • Depth: 18.1 In (46 cm)
    • Weight: < 20 lbs (< 9.07 kg)
  • General:
    • Contact Monitoring Systems: BovieNEM
    • Range/Trip Point: 10 ohms or less over entire operating range (10C° to 40°C)
    • Cooling System: Passive Cooling
    • Technology: Isolated RF Output
    • Pre-set Feature: Yes
    • Monopolar HP Ports: 2
    • Dual Activation Tone Alert: Yes
    • Connector Ports/Power Switch: White
    • Power Display Accuracy: ± 20% or 5 watts whichever is greater
  • Bipolar:
    • Macro Bipolar:
      • Watts: 80
      • VPeak Max Voltage: 600V
      • Crest Factor (Typical): 1.5 ± 20%
      • Frequency: 490 kHz ± 4.9 kHz
    • Micro Bipolar:
      • Watts: 80
      • VPeak Max Voltage: 500V
      • Crest Factor (Typical): 1.5 ± 20%
      • Frequency: 490 kHz ± 4.9 kHz
    • Standard Bipolar:
      • Watts: 50
      • VPeak Max Voltage: 250V
      • Crest Factor (Typical): 1.5 ± 20%
      • Frequency: 490 kHz ± 4.9 kHz
    • Monopolar:
      • Cut I:
        • Watts: 200
        • VPeak Max Voltage: 1000V
        • Crest Factor (Typical): 1.7 ± 20%
        • Frequency: 490 kHz ± 4.9 kHz
      • Cut II:
        • Watts: 200
        • VPeak Max Voltage: 750V
        • Crest Factor (Typical): 1.7 ± 20%
        • Frequency: 490 kHz ± 4.9 kHz
      • Blend (1) – Blend (4):
        • Watts: 200
        • VPeak Max Voltage: (Varies)
        • Crest Factor (Typical): (Varies)
        • Frequency: 490 kHz ± 4.9 kHz
      • Pinpoint Coagulation:
        • Watts: 120
        • VPeak Max Voltage: 1800V
        • Crest Factor (Typical): 3.1 ± 20%
        • Frequency: 490 kHz ± 4.9 kHz
      • Spray Coagulation:
        • Watts: 120
        • VPeak Max Voltage: 4000V
        • Crest Factor (Typical): 6.0 ± 20%
        • Frequency: 350 to 450 kHz
      • Gentle Coagulation:
        • Watts: 120
        • VPeak Max Voltage: 450V
        • Crest Factor (Typical): 1.6 ± 20%
        • Frequency: 490 kHz ± 4.9 kHz