Bovie IDS 400

Single Product Request Form
Category

Description

DESCRIPTION: Bovie IDS 400 Features

  • The Bovie IDS 400 integrates IDS technology, offering surgeons flexibility, convenience, and safety in choosing the desired effect consistently.
  • Innovative electrosurgical generator with fully digital implementation for use in modern OR and surgical outpatient centers.
  • Capable of covering a spectrum of electrosurgical interventions with monopolar and bipolar functions, ensuring maximum safety and user-friendly interface.
  • BovieFDFS (Fast Digital Feedback System): Digitally controlled CUT and COAG functions provide consistent clinical output.
  • BovieNEM (Neutral Electrode Monitoring).
  • BovieFCFS (First Come First Serve): Improved patient safety with the latest digital implementation.
  • BovieDED (Digital Error Detection).

SPECIFICATIONS: Bovie IDS 400 Specifications

  • Dimensions:
    • Height: 6.5 In (16.5 cm)
    • Width: 14.75 In (37.5 cm)
    • Depth: 18.1 In (46 cm)
    • Weight: < 20 lbs (< 9.07 kg)
  • General:
    • Duty Cycle: 10 S / 30 S
    • Output Type: RF Isolated “F”
    • Output Frequency: 490 kHz ± 5
    • Input Current: 4.5 A ~
    • Line Voltage: 100-240 V ~
  • Operating Parameters:
    • Ambient temperature range: 10° to 40° C
    • Relative humidity: 30% to 75%, non-condensing
    • Atmospheric pressure: 70kPa to 106kPa
    • Warm-up time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.
  • Transport and Storage:
    • Ambient temperature range: -40° to +70° C
    • Relative humidity: 10% to 100%, including condensation
    • Atmospheric pressure: 50kPa to 106kPa
  • Modes:
    • Cut I:
      • Output Power: 400 W @ 500Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Crest Factor at Rated Load: 1.6 ± 20%
      • VPeak Max: 1000V
    • Cut II:
      • Output Power: 300 W @ 300Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Crest Factor at Rated Load: 1.6 ± 20%
      • VPeak Max: 625V
    • Blend (Max):
      • Output Power: 200 W @ 300Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Repetition Rate: 30 kHz ± 5 kHz
      • Crest Factor at Rated Load: 3.5 ± 20%
      • VPeak Max: 2000V
    • Pinpoint:
      • Output Power: 120 W @ 500Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Repetition Rate: 30 kHz ± 5 kHz
      • Crest Factor at Rated Load: 4.5 ± 20%
      • VPeak Max: 2400V
    • Spray:
      • Output Power: 80 W @ 500Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Repetition Rate: 30 kHz ± 5 kHz
      • Crest Factor at Rated Load: 6.5 ± 20%
      • VPeak Max: 4000V
    • Bipolar:
      • Output Power: 120 W @ 150Ω
      • Output Frequency: 490 kHz ± 4.9 kHz
      • Crest Factor at Rated Load: 1.6 ± 20%
      • VPeak Max: 450V

Top of Form