ConMed Excalibur Plus

Single Product Request Form
Category

Description

ConMed Excalibur Plus Features:

  • Most Powerful Isolated ESU: The ConMed Excalibur Plus PC is the most potent isolated Electrosurgical Unit (ESU) available in its class, providing exceptional performance and versatility in surgical procedures.
  • Adjustable Power Settings: Surgeons can conveniently adjust power settings from the surgical field using a standard ES pencil, enhancing control and efficiency during procedures.
  • Patented Controlled Output Voltage in CUT: The Excalibur Plus PC features patented technology that maintains the minimum necessary voltage for smooth cutting through various impedance tissues, minimizing thermal damage and ensuring precision.
  • Aspen Return Monitor (ARMU): Incorporates the Aspen Return Monitor (ARMU), the most sensitive pad monitoring system available, which combines automatic feedback circuitry and visual displays to provide precise contact quality monitoring during surgeries.
  • Electrosurgical Generator with Multiple Modes: Offers Cut, Coag, Bipolar, and Blend modes, providing flexibility and adaptability for various surgical applications.
  • Output Power: 300 watts of output power, ensuring sufficient energy delivery for demanding procedures.
  • Scalpel-like Cut: The Excalibur Plus PC delivers a scalpel-like cut and offers three blend modes for enhanced surgical precision and control.
  • Peak Voltage Specifications:
    • Pure Cut: Maximum Peak to Peak Voltage of 2000 Volts
    • Blend 1: Maximum Peak to Peak Voltage of 2000 Volts
    • Blend 2: Peak Voltage of 2200 Volts
    • Blend 3: Maximum Peak to Peak Voltage of 2400 Volts

SPECIFICATIONS: ConMed Excalibur Plus

  • Dimensions:
    • Height: 6.25 in (15.9 cm)
    • Width: 18 in (45.8 cm)
    • Length: 22.75 in (57.8 cm)
    • Weight: 40 lbs. (18.1 kg)
  • Environmental Conditions:
    • Operating Conditions: 10°C to 30°C, 95% RH non-condensing
    • Storage Conditions: -14°C to 70°C, altitude -60 to +4500 m MSL when sealed in original poly bag
  • Line Regulation: Power change < 1% /V for vline = min to max voltage
  • Power Display Accuracy: The greater of +/- 10% of display or +/- 3 watts rated load.
  • Pad Monitor:
    • Single Pad Mode: Two-wire continuity detector, typical threshold = 10 Ohms
    • Dual Pad Mode: Two-wire resistance monitor, typical acceptance range 10 to 150 OHMS, trip threshold typically 20% higher than the last monitor set point activation, visual indication of patient resistance.
  • Activation Frequencies:
    • Cut: 520 HZ
    • COAG: 400 HZ
    • Operator Error: 1710 HZ, pulsating
    • Plate Alarm: 1000 HZ (65 DBA MINIMUM)
    • Machine Fault: 1000 HZ (45 DBA Min. Except Alarms Are 65 DBA Min)