  1. Central Processing Unit: The GE Marquette Solar 8000 Patient Monitor serves as the processing unit for the Solar 8000 monitoring system. It acts as the central hub, providing user controls and communication with various patient monitoring modules housed in the Tram-rac. These modules analyze patient data, allowing for comprehensive monitoring.
  2. Compatibility: The processing unit is compatible with many display units, allowing for flexible configuration based on the needs of the clinical setting. It can interface with different display options to present patient data effectively.
  3. Multi-Waveform Display: The Solar 8000 system can display up to 8 waveforms simultaneously on a monitor, providing clinicians with comprehensive real-time data for patient monitoring and assessment.
  4. High-Performance Monitor: The Solar 8000 system is designed to deliver high-performance monitoring capabilities, ensuring accurate and reliable patient data acquisition and analysis.
  5. Parameters Monitored: The system monitors essential patient parameters, including ECG (Electrocardiography), NIBP (Non-Invasive Blood Pressure), SpO2 (Oxygen Saturation), and Temperature. It also supports monitoring of two IBP (Invasive Blood Pressure) channels.
  6. Options: The system offers optional features such as a flat screen display for enhanced user experience and flexibility in monitoring.

GE Marquette Solar 8000 Specifications:

  • Dimensions:
    • Height: 8.1 cm (3.2 in)
    • Width: 33.6 cm (13.1 in)
    • Depth: 34.9 cm (13.7 in)
    • Weight: 5.4 kg (12 lbs)
  • Controls:
    • Trim Knob and five active hard keys: Display On/Off, NBP Go/Stop, Zero All, Silence Alarm, and Graph Go/Stop
    • Optional remote control with twelve active hard keys
  • Processing:
    • Main processor: Motorola® MC68EN360, 32 bits, 25 MHz
    • Graphics processor: Texas Instruments® TMS34010, 16 bits, 46.8 MHz
    • Communication processors: Intel® 82596, Motorola MC68EN360
  • Alarms:
    • Classification: Patient Status Alarms (Crisis, Warning, Advisory, and Message), System Alarms (Warning and Advisory)
    • Notification: Audible and visual
    • Display of alarm information: All limits are viewable and graphable, continuous display of limits for all parameters
  • Power Requirements:
    • 110 ± 20VAC, 50/60Hz single phase or 230 ± 40VAC, 50/60Hz single phase
    • Power consumption: 100 W (max.)
    • Low voltage shutdown: < 90 VAC
    • Cooling: Natural convection
    • Heat dissipation: 100 Btu/hr (30 watts)
  • Operating Conditions:
    • Ambient Temperature: 10°C to 40°C (50°F to 104°F)
    • Relative Humidity: 10% to 90% (non-condensing)
  • Storage Conditions:
    • Temperature: -40°C to 70°C (-40°F to 158°F)
    • Relative Humidity: 15% to 95% (non-condensing)