Mindray V21

Single Product Request Form
Category

Description

Features:

  1. Modular Patient Monitoring System: The Mindray V21 monitoring platform offers a modular architecture, allowing customization of monitored parameters based on patients’ needs or acuity levels through the addition or removal of modules.
  2. Portrait Display Orientation: Maximizing the amount of viewable data, the V21 utilizes a portrait display orientation, providing clear visibility of up to 12 waveforms simultaneously on its 21.3-inch color touchscreen display.
  3. Integrated Connectivity: The V21 can integrate with Panorama Central Stations and Nurse Call Systems, enhancing communication and workflow efficiency in clinical settings.
  4. IV Drug and Hemodynamic Calculations: This monitoring platform supports IV drug calculations and hemodynamic calculations, aiding clinicians in medication administration and patient management.
  5. Modern User Interface: With a modern user interface, the V21 ensures ease of use and efficient operation for healthcare professionals.
  6. Interchangeable System Elements: The V21 features interchangeable system elements, providing flexibility and adaptability to various clinical environments.
  7. Consistent Performance and Easy Maintenance: Designed for consistent performance and easy maintenance, the V21 platform ensures reliability and longevity in clinical use.

Mindray V21 Specifications:

  • Dimensions:
    • Height: 23 in (58.4 cm)
    • Width: 16 in (40.6 cm)
    • Depth: 9 in (22.9 cm)
    • Weight: 23 lbs (10.4 kg) (Without batteries)
  • Display:
    • Type: TFT Display
    • Size: 21.3 in (54.1 cm)
    • Aspect Ratio: 3:4
    • Resolution: 1024 x 768 pixels
  • ECG (3, 5, 6 Lead):
    • Range: Neonatal BPM: 30 to 350; Pediatric and Adult BPM: 30 to 300
    • Resolution: 1 BPM
    • Accuracy: ±3 BPM or ±3% from 30 to 250 BPM, ±5% in the 251 to 350 BPM range
  • ECG (12 Lead):
    • Range: Neonatal, Pediatric, and Adult BPM: 30 to 300
    • Resolution: 1 BPM
    • Accuracy: ±3 BPM or ±3% from 30 to 250 BPM, ±5% in the 251 to 350 BPM range
  • NIBP Systolic/Diastolic Pressure:
    • Accuracy: Mean error ±5 mmHg
    • Standard Deviation: Less than 8 mmHg
    • Measurement Range: Varies based on mode (Adult, Pediatric, Neonate)
  • NIBP Heart Rate:
    • Accuracy: Greater of ±3 BPM or ±3%
    • Measurement Range: Varies based on mode (Adult, Pediatric, Neonate)
  • IBP:
    • Accuracy: ±2mmHg or 2% of the signal value, whichever is greater
    • Resolution: 1 mmHg
  • Temperature:
    • Range: 59°F to 113°F (15°C to 45°C)
    • Resolution: 0.1°F (0.1°C)
    • Accuracy: ±0.2°F (±0.1°C)
  • SpO2:
    • Range: 70% to 100%
    • Resolution: 1%
    • Update Rate: 2 seconds
  • Thermomodulation Cardiac Output (CO):
    • Range: 0.2 to 20.0 liters/minute
    • Resolution: 0.1 liters/minute
    • Accuracy: 5% or 0.2 liters/minute, whichever is greater
  • CO2:
    • Performance: 0-99 mmHg
    • Respiration Rate Range: 0 to 150 RPM
    • Flow Rate: 50 ml/min -7.5 + 15 ml/min