Description
Stryker Smart Pump Features:
- Supports simultaneous surgery on both upper and lower limbs with its dual-channel design.
- Each cuff’s pressure and time settings are displayed, controlled, and monitored independently.
- Procedure timer and cuff target pressure can be easily set using intuitive steps.
- Control buttons are clearly identified for user convenience.
Stryker Smart Pump Specifications:
- Power Sources:
- AC Input: 100 – 240 VAC, 50/60 Hz universal switching power supply.
- Automatic internal battery switchover for AC-free use or AC loss protection.
- Facility power: 120 VAC, 10A, hospital-grade receptacle.
- Internal DC Supply Module:
- Output voltage: 15 VDC ± 1%, 150 mV pk-pk ripple.
- Output current: 4A maximum.
- Output power: 60 W maximum.
- Internal Battery:
- Type: Nickel metal hydride (NiMH).
- Nominal voltage: 12 VDC.
- Rated capacity: 2.6 Ahr.
- Charge time: 1 hour (80% charge), 6 hours (full charge).
- Power conservation: 30 days (fully charged, then unplugged).
- Display:
- Type: High-resolution backlit custom LCD.
- Timer: Elapsed time numeric.
- Pressure: Real-time digital numeric with graphic pressure gauge.
- Air Pump:
- Nominal power: 12 VDC, 1.25 A.
- Flow: 10 LPM open flow.
- Pressure Transducers:
- Operating pressure: 0 – 15 psi.
- Maximum pressure: 30 psi.
- Repeatability: ± 0.2% FS typical.
- Cuff Pressure:
- Accuracy: ± 2% or 4 mmHg, whichever is greater, of set pressure value.
- Range: 100 to 475 mmHg.
- Cuff Timer:
- Range: 1 to 240 minutes.
- Adjustment: 1 to 15 minutes (1-minute increments); 15 to 240 minutes (5-minute increments).